FDA Medical Device Class II: Baxter Healthcare Corporation – TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Mode
Class II issued for TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087. There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use....
The FDA Center for Devices and Radiological Health (CDRH) has issued an official medical device recall determination for TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087 produced by Baxter Healthcare Corporation.
Regulatory Recall Specification
- Recall Number: Z-0354-2027
- Recall Classification: Class II
- Firm Name: Baxter Healthcare Corporation
- Current Status: Ongoing
Reason for Device Recall
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Clinical & Corrective Actions
Healthcare facilities and clinical engineering departments are advised to review the affected serial numbers and follow the manufacturer's field safety notice instructions.